MedElevator Войти в исследования
← Поиск публикаций

DOI 10.12659/msm.896085

Медицинская библиотека · Medical science monitor : international medical journal of experimental and clinical research · 2015

Biomechanics of Hybrid Anterior Cervical Fusion and Artificial Disc Replacement in 3-Level Constructs: An In Vitro Investigation.

Liao Zhenhua, Fogel Guy R, Pu Ting, Gu Hongsheng, Liu Weiqiang

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The ideal surgical approach for cervical disk disease remains controversial, especially for multilevel cervical disease. The purpose of this study was to investigate the biomechanics of the cervical spine after 3-level hybrid surgery compared with 3-level anterior cervical discectomy and fusion (ACDF).

MATERIAL AND

METHODS

Eighteen human cadaveric spines (C2-T1) were evaluated under displacement-input protocol. After intact testing, a simulated hybrid construct or fusion construct was created between C3 to C6 and tested in the following 3 conditions: 3-level disc plate disc (3DPD), 3-level plate disc plate (3PDP), and 3-level plate (3P).

RESULTS

Compared to intact, almost 65~80% of motion was successfully restricted at C3-C6 fusion levels (p<0.05). 3DPD construct resulted in slight increase at the 3 instrumented levels (p>0.05). 3PDP construct resulted in significant decrease of ROM at C3-C6 levels less than 3P (p<0.05). Both 3DPD and 3PDP caused significant reduction of ROM at the arthrodesis level and produced motion increase at the arthroplasty level. For adjacent levels, 3P resulted in markedly increased contribution of both upper and lower adjacent levels (p<0.05). Significant motion increases lower than 3P were only noted at partly adjacent levels in some conditions for 3DPD and 3PDP (p<0.05).

CONCLUSIONS

ACDF eliminated motion within the construct and greatly increased adjacent motion. Artificial cervical disc replacement normalized motion of its segment and adjacent segments. While hybrid conditions failed to restore normal motion within the construct, they significantly normalized motion in adjacent segments compared with the 3-level ACDF condition. The artificial disc in 3-level constructs has biomechanical advantages compared to fusion in normalizing motion.

Для цитирования: Liao Zhenhua, Fogel Guy R, Pu Ting, Gu Hongsheng, Liu Weiqiang. Biomechanics of Hybrid Anterior Cervical Fusion and Artificial Disc Replacement in 3-Level Constructs: An In Vitro Investigation.. Medical science monitor : international medical journal of experimental and clinical research. 2015. DOI: 10.12659/msm.896085.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.