MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.spinee.2012.07.002

Медицинская библиотека · The spine journal : official journal of the North American Spine Society · 2012

A biomechanical evaluation of a spacer with integrated plate for treating adjacent-level disease in the subaxial cervical spine.

Beutler William J, Clavenna Andrew L, Gudipally Manasa, Moldavsky Mark, Khalil Saif

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Adjacent level degeneration (ALD) has been reported as one of the long-term consequences of anterior discectomy and fusion despite its clinical success in treating cervical pathologies. Traditionally, ALD is treated by replacing the previously implanted plate with a longer plate, which can lead to postoperative complications. The biomechanics of SIP in the adjacent level has not been investigated.

PURPOSE: To evaluate the multidirectional stability of a spacer with integrated plate (SIP) in comparison to a traditional spacer and plate (TSP).

STUDY

DESIGN

To evaluate the biomechanical stability of a spacer with integrated plate adjacent to a traditional spacer and plate construct in a human cervical cadaveric model.

METHODS

Eight fresh human cervical (C2-C7) cadaver spines were mounted on a six degree-of-freedom spine simulator. The sequence of test constructs was: 1) Intact; 2) TSP (C4-C6) with SIP (C3-C4); and 3) TSP (C3-C6). An unconstrained moment of ±1.5 Nm was used in flexion-extension, lateral bending, and axial rotation. Range of motion (ROM) was measured by a digital motion analysis system. Statistical analysis was performed using ANOVA repeated measures.

RESULTS

All instrumented constructs significantly reduced ROM compared to the intact condition. No statistically significant difference was observed between the two-level TSP with an adjacent SIP construct and three-level TSP construct in all loading modes.

CONCLUSION

The biomechanical study shows that adding a spacer with integrated plate adjacent to a two-level anterior plate demonstrates equivalent stability to a three-level anterior plate. The spacer with integrated plate, which preserves the originally plated fusion levels, may overcome the complications associated with the traditional technique of replacing the original plate with a longer plate. However, prospective clinical studies are required to address the clinical benefits and challenges, if any.

Для цитирования: Beutler William J, Clavenna Andrew L, Gudipally Manasa, Moldavsky Mark, Khalil Saif. A biomechanical evaluation of a spacer with integrated plate for treating adjacent-level disease in the subaxial cervical spine.. The spine journal : official journal of the North American Spine Society. 2012. DOI: 10.1016/j.spinee.2012.07.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.