STUDY

DESIGN

Systematic review with meta-analysis.

OBJECTIVE

This meta-analysis aims to assess robotic and conventional surgical techniques for instrumentation of the spine among pediatric, adolescent, and young adult populations.

SUMMARY OF BACKGROUND DATA: Spinal deformities among younger patients may be treated with pedicle screw instrumentation; however, freehand guidance often leads to screw misplacements and consequent complications. Although computer-navigation surgery improved the screw accuracy, physiological limitations of the surgeon still remain a serious challenge. Robot-assisted (RA) surgery became a novel technique for enhancing screw accuracy.

METHODS

PRISMA and Cochrane Handbook were followed, and five databases were searched from inception. Studies analyzing patients aged up to 25 years old, and comparing RA to freehand or navigation were included. Outcomes included screw accuracy, perioperative outcomes, and complications. Meta-analysis was pooled with random-effects model.

RESULTS

Finally, 10 studies, 550 patients, and 8061 screws were included in the meta-analysis. Accurate screw placement rates (grade 0: OR 2.33, P <0.001) and clinically acceptable placements (grade 0+1: OR 3.09, P <0.001) were significantly higher in robotic cohort. However, RA surgery increased operation time (MD 21.51 min, P =0.03). Blood loss, Cobb angle correction, hospital stay, or complications showed no significant differences between the two groups.

CONCLUSIONS

RA surgery leads to improved accuracy of the pedicle scores; however, this does not directly translate to better clinical or curvature outcomes. Moreover, longer operation times were observed in the robotic cohort. More high-quality studies are needed to validate these findings.

Для цитирования: Łajczak Anna, Łajczak Paweł, Jóźwik Kamil, Jaldin Torrico Cristian, Nowakowski Przemysław, Buczkowski Stanisław, Ayesha Ayesha. Robot-assisted Versus Conventional Pedicle Screw Instrumentation for Pediatric, Adolescent, and Young Adult Populations-Meta-analysis.. Spine. 2025. DOI: 10.1097/brs.0000000000005488.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.