DOI 10.1177/23094990211042237

Медицинская библиотека · Journal of orthopaedic surgery (Hong Kong) · 2021

A biomechanical study on the effect of lengthening magnitude on spine off-loading in magnetically controlled growing rod surgery: Implications on lengthening frequency.

Wong Douglas C, Nafo Wanis, Lu William Weijia, Chee Cheung Kenneth Man

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Purpose: To assess whether the magnitude of lengthening in magnetically controlled growing rod (MCGR) surgeries has an immediate or delayed effect on spinal off-loading.

Methods

9 whole porcine spines were instrumented using two standard MCGRs from T9 to L5. Static compression testing using a mechanical testing system (MTS) was performed at three MCGR lengthening stages (0 mm, 2 mm, and 6 mm) in each spine. At each stage, five cycles of compression at 175N with 25 min of relaxation was carried out. Off-loading was derived by comparing the load sustained by the spine with force applied by the MTS to the spine. Micro-CT imaging was subsequently performed.

Results

The mean load sustained by the vertebral body before lengthening was 39.69N, and immediately after lengthening was 25.12N and 19.91N at 2 mm and 6 mm lengthening, respectively; decreasing to 10.07N, 8.31N, and 8.17N after 25 minutes of relaxation, at 0 mm, 2 mm, and 6 mm lengthening stages, respectively. There was no significant difference in off-loading between 2 mm and 6 mm lengthening stages, either instantaneously (p = 0.395) or after viscoelastic relaxation (p = 0.958). CT images showed fractures/separations at the level of pedicle screws in six spines and in the vertebral body's growth zone in five spines after 6 mm MCGR lengthening.

Conclusion

This study demonstrated MCGRs cause significant off-loading of the spine leading to stress shielding. 6 mm of lengthening caused tissue damage and microfractures in some spines. There was no significant difference in spine off-loading between 2 mm and 6 mm MCGR lengthening, either immediately after lengthening or after viscoelastic relaxation.

Для цитирования: Wong Douglas C, Nafo Wanis, Lu William Weijia, Chee Cheung Kenneth Man. A biomechanical study on the effect of lengthening magnitude on spine off-loading in magnetically controlled growing rod surgery: Implications on lengthening frequency.. Journal of orthopaedic surgery (Hong Kong). 2021. DOI: 10.1177/23094990211042237.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.