Weight-bearing recommendations after first metatarsophalangeal joint arthrodesis fixation: a biomechanical comparison.

Campbell Bradley, Schimoler Patrick, Belagaje Sudhir, Miller M C, Conti S F

СтатьяПеревод в очереди

Аннотация

BACKGROUND

This study sought to determine whether several metatarsophalangeal (MTP) fusion techniques require complete immobilization or if some level of weight-bearing could be recommended after surgery. A comparison of synthetic composite to actual bone was included in order to examine the validity of the testing conditions.

METHODS

Four MTP fusion modalities were tested in synthetic composite bone models: unlocked plating, locked plating, crossed lag screws, and an unlocked plate with a single lag screw. Stiffness was calculated and then used to find the two most rigid constructs; the load to failure was recorded. Stiffness and load to failure testing for the two more rigid constructs in paired cadaveric bones were followed.

RESULTS

The unlocked plate plus screw and crossed screw constructs were stiffest (p < 0.008). Loads to failure of the unlocked plate plus screw and crossed screws in synthetic bone were 131 and 101 N, respectively and in cadaveric bone were 154 and 94 N, respectively, which are less than the estimated 25% body weight required at the MTP joint. The plate plus screws were statistically more stiff than crossed screws (p = 0.008), but there was no statistical difference between synthetic and cadaveric bone in load to failure (p = 0.296).

CONCLUSIONS

The plate plus screw offered the greatest stiffness; the failure test showed that no construct could withstand weight-bearing as tolerated; and, synthetic composite models of the MTP joint did not provide the consistent results in stiffness and failure.

Для цитирования: Campbell Bradley, Schimoler Patrick, Belagaje Sudhir, Miller M C, Conti S F. Weight-bearing recommendations after first metatarsophalangeal joint arthrodesis fixation: a biomechanical comparison.. Journal of orthopaedic surgery and research. 2017. DOI: 10.1186/s13018-017-0525-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.