Mixed reality guidance in osteotomy provides superior precision and accuracy: validation and comparative study.

Stefaniak Jakub, Stupnicki Szymon, Tuczyński Piotr, Kaczmarek Agata, Niziński Jan, Rajczyk Justyna, Lubiatowski Bartłomiej, Lubiatowski Przemysław

СтатьяСравнительное исследованиеВалидационное исследованиеПеревод в очереди

Аннотация

PURPOSE: Bone deformities, such as cubitus varus, can lead to abnormal joint alignment and impaired function. Corrective osteotomy aims to restore anatomical alignment, and its precision may be enhanced using various guidance methods. Emerging mixed reality systems allow for the placement and manipulation of virtual objects and may offer effective surgical navigation. This study aimed to validate the accuracy and precision of holographic guidance compared with classic visual estimation and a printed triangle gauge in a controlled laboratory setting.

METHODS

Closed-wedge osteotomies at angles of 15° and 30° were performed on fresh-frozen porcine femora. Three techniques were evaluated: Group EB (eyeballing)-visual estimation; Group PW (printed wedge)-using 15° and 30° plastic templates; and Group HW (holographic wedge)-using a mixed reality system (RSQ HOLO, RSQ Technologies, Poznań, Poland, and HoloLens 2, Microsoft). In the HW group, a holographic wedge tool guided the osteotomy. The angle of the excised bone wedge and post-osteotomy alignment in the anteroposterior (AP) and lateral planes were measured physically (goniometer) and digitally (radiographs). Statistical analysis assessed accuracy (closeness to 15°/30°) and precision (standard deviation [SD], mean deviation, coefficient of variation [CV]).

RESULTS

At 15°, the PW and HW techniques demonstrated greater accuracy and lower error compared with EB (T = 3.60; p < 0.01), with HW yielding the lowest systematic error (SE = 0.12). Alignment in the AP plane was similar across groups (T < 2.11; p > 0.05), whereas lateral alignment showed significant differences (T > 2.11; p < 0.05). At 30°, HW achieved the best alignment in both AP (T = 2.48; p = 0.01) and lateral views (T = 2.35; p = 0.02). Lateral alignment was improved by more than 5° with holographic guidance compared to other techniques (p = 0.01). The HW group exhibited the highest precision for both 15° and 30° angles (lowest SD and CV; p < 0.05).

CONCLUSIONS

Augmented reality provides accurate and precise intraoperative guidance, outperforming both visual estimation and printed wedge templates in deformity correction.

Для цитирования: Stefaniak Jakub, Stupnicki Szymon, Tuczyński Piotr, Kaczmarek Agata, Niziński Jan, Rajczyk Justyna, Lubiatowski Bartłomiej, Lubiatowski Przemysław. Mixed reality guidance in osteotomy provides superior precision and accuracy: validation and comparative study.. International orthopaedics. 2025. DOI: 10.1007/s00264-025-06574-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.