A Novel Algorithm Using Within-leg Calibration for Enhanced Accuracy of Detection of Arthritis by Infrared Thermal Imaging in Children.
Аннотация
OBJECTIVE
To standardize and improve the accuracy of detection of arthritis by thermal imaging.
METHODS
Children with clinically active arthritis in the knee or ankle, as well as healthy controls, were enrolled to the development cohort; another group of children with knee symptoms was enrolled to the validation cohort. Ultrasound was performed in the arthritis subgroup for the development cohort. Joint exam by certified rheumatologists was used as a reference for the validation cohort. Infrared thermal data were analyzed using custom software. Temperature after within-limb calibration (TAWiC) was defined as the temperature differences between joint and ipsilateral mid-tibia. TAWiC of knees and ankles was evaluated using ANOVA across subgroups. Optimal thresholds were determined by receiver-operating characteristic analysis using Youden index.
RESULTS
There were significant differences in mean and 95th TAWiC of knee in anterior, medial, lateral views, and of ankles in anterior view, between inflamed and uninflamed counterparts (P < 0.05). The area under the curve was higher by 30% when using TAWiCknee than that when using absolute temperature. Within the validation cohort, the sensitivity of accurate detection of arthritis in the knees using both mean and 95th TAWiC from individual views or all 3 views combined ranged from 0.60 to 0.70, and the specificity was > 0.90 in all views.
CONCLUSION
Children with active arthritis or tenosynovitis in knees or ankles exhibited higher TAWiC than healthy joints. Our validation cohort study showed promise for the clinical utility of infrared thermal imaging for arthritis detection.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.