Biomechanical analysis of a magnesium plantar plate prototype system for the first tarsometatarsal joint fusion: a cadaveric study.

Zhou Peng, Ribeiro Marx, Greven Johannes, Praster Maximilian, Seitz Jan-Marten, Habicht Simon, Hildebrand Frank, Balmayor Elizabeth R, Lichte Philipp

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Titanium plantar plates have proven successful in the fixation of the first tarsometatarsal arthrodesis (TMT). However, a second surgery is typically needed for implant removal, and potential adverse effects, carried by the conventional implantations, are not uncommon. The purpose of this study was to determine whether a novel magnesium-based plantar plate system provides similar fusion stability to a titanium-based plantar plate system under various loading conditions.

METHODS

Six matched-pair human cadaveric specimens underwent TMT fusions using either a magnesium plantar plate system prototype or a titanium plantar plate system. Specimens were cyclically loaded with a force ranging from 5 N to 50 N for 5,000 cycles, and displacement was recorded. Axial stiffness (N/mm) was calculated from load-displacement curves. Each specimen was loaded to failure at a rate of 5 mm/min, and the ultimate load was recorded.

RESULTS

No significant difference was found in the vertical displacement between Ti group and Mg group after 100 cycles (2.4 ± 1.0 mm vs. 1.3 ± 1.4 mm, p = 0.196), 500 cycles (3.3 ± 1.3 mm vs. 1.7 ± 1.7 mm, p = 0.142), 1,000 cycles (3.7 ± 1.5 mm vs. 1.9 ± 1.9 mm, p = 0.128), 2,500 cycles (4.2 ± 1.7 mm vs. 2.3 ± 2.2 mm, p = 0.172) and 5,000 cycles (4.5 ± 1.8 mm vs. 2.3 ± 3.3 mm, p = 0.125), Additionally, no significant differences were observed in initial stiffness (53.1 ± 19.2 N/mm vs. 82.2 ± 53.9 N/mm, p = 0.257), final stiffness (90.6 ± 48.9 N/mm vs. 120.0 ± 48.3 N/mm, p = 0.319), or maximum load-to-failure (259.8 ± 98.2 N vs. 323.9 ± 134.9 N, p = 0.369).

CONCLUSIONS

Based on the performed biomechanical testing, the magnesium plantar plate system provides mechanical stability equivalent to the titanium plantar plate system in fixation for the first TMT joint fusion.

Для цитирования: Zhou Peng, Ribeiro Marx, Greven Johannes, Praster Maximilian, Seitz Jan-Marten, Habicht Simon, Hildebrand Frank, Balmayor Elizabeth R, Lichte Philipp. Biomechanical analysis of a magnesium plantar plate prototype system for the first tarsometatarsal joint fusion: a cadaveric study.. Journal of orthopaedic surgery and research. 2024. DOI: 10.1186/s13018-024-05208-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.