STUDY

DESIGN

Retrospective observational study.

OBJECTIVE

To scrutinize screw motion used in semiconstrained rotational plate systems for anterior cervical discectomy and fusion (ACDF).

SUMMARY OF BACKGROUND DATA: Semiconstrained rotational plate systems are supposed to control graft subsidence and facilitate lordosis acquisition and maintenance by toggling the instrumented vertebrae via variable-angle screws. However, their benefits may be unrealized if the screws move within the vertebrae.

METHODS

We reviewed medical records of 119 patients who underwent 1-level, 2-level, 3-level, or 4-level ACDF, divided them into the short-segment (n=62, 1-level or 2-level ACDF) and long-segment (n=59, 3- level or 4-level ACDF) groups, and investigated their immediate and 1-year postoperative lateral radiographs. We measured the fused segmental angle, screw angles at the upper-instrumented vertebra (UIV) and lower-instrumented vertebra (LIV), distance from the screw base to the endplate of UIV/LIV (SBE), and distance from the screw tip to the endplate of UIV/LIV (STE) to analyze the screw motion used in these plate systems. The differences between the immediate and 1-year postoperative values were statistically analyzed. The nonunion level was also investigated.

RESULTS

Screw angle and SBE at the LIV significantly decreased in the long-segment group (-14.5±9.8 degrees and -2.8±1.8 mm, respectively) compared with those in the short-segment group (-4.6±6.0 degrees and -1.0±1.5 mm, respectively). Thus, the long-segment group could not maintain the immediate-postoperative segmental angle. Overall, 27 patients developed nonunion, with 19 (70.4%) in the long-segment group and 21 (77.8%) at the lowest fused level.

CONCLUSIONS

Semiconstrained rotational plate systems provide only vertical forces to the fused segment rather than toggling the instrumented vertebrae. Postoperatively in multilevel ACDF, LIV screws migrate caudally, suggesting that these plate systems are not always effective in maintaining lordosis. Moreover, LIV screws and the anterior wall of the LIV are subject to overloading, resulting in a high rate of nonunion at the lowest fused level.

LEVEL OF EVIDENCE: Level III.

Для цитирования: Tatara Yasunori, Niimura Takanori, Sakaguchi Akira, Katayama Hiroki, Miyaoka Yoshinari, Mihara Hisanori. Screw Motion Used in Semiconstrained Rotational Plate Systems for Anterior Cervical Discectomy and Fusion.. Clinical spine surgery. 2024. DOI: 10.1097/bsd.0000000000001665.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.