m+Medelevator.RU
← Поиск публикаций
Joint diseases and related surgery · 2026

Optimizing fixation in medial distal femur fractures: A biomechanical study.

СтатьяСравнительное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Durgal Atahan, Kilic Enver, Perdeci Elif Naz, Erbay Elibol Fatma Kubra, Karlidag Taner, Kulakoglu Burak, Demir Teyfik, Bingol Olgun, Ozdemir Guzelali

Аннотация

OBJECTIVES: This study aims to biomechanically compare four fixation methods to identify the most stable construct under axial loading conditions.

MATERIAL AND METHODS: A standardized osteotomy model simulating AO/OTA 33-B2.1 fractures was created in 28 synthetic femurs. Four groups (n = 7 each) were tested: (1) three 6.5-mm cannulated screws, (2) the same with an additional centrally placed screw, (3) three screws plus a 3.5-mm buttress plate, and (4) three screws combined with a reversed anatomical proximal tibial locking plate. Constructs were subjected to cyclic and static axial loading. The prespecified primary endpoint was load at 2-mm displacement, with stiffness and maximum load to failure as secondary outcomes.

RESULTS: Group 4 exhibited the highest stiffness (802 ± 70 N/mm), maximum load (2360 ± 389 N), and resistance to 2-mm displacement (1138 ± 87 N). Group 2, with a centrally placed fourth screw, showed significantly improved performance in terms of maximum load compared to the traditional three-screw construct. While Group 3 improved stiffness over screw-only constructs, it did not significantly outperform Group 2 in key load metrics. Group 2, with a centrally placed fourth screw, demonstrated comparable stability to the buttress plate and anatomical locking plate in selected metrics, but remained inferior to the locking plate construct.

CONCLUSION: Central placement of cannulated screws to the medial femoral condyle enhanced axial stability in this synthetic model and represented a minimally invasive alternative to buttress plating in selected metrics. The highest stability was achieved with the combined use of cannulated screws and a locking plate. These results are restricted to axial loading and the tested endpoints; thus, further biomechanical, cadaveric, and clinical studies are warranted before generalizing to early weight-bearing. Given the anatomical challenges of the medial condyle, the development of anatomically contoured locking plates specifically designed for this region remains an important future direction.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты