World neurosurgery · 2021

A Biomechanical Model for Testing Cage Subsidence in Spine Specimens with Osteopenia or Osteoporosis Under Permanent Maximum Load.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Schleifenbaum Stefan, Heilmann Robin, Riemer Elena, Reise Rebekka, Heyde Christoph-Eckhard, Jarvers Jan-Sven, Pieroh Philipp, Völker Anna, von der Hoeh Nicolas Heinz

Аннотация

BACKGROUND: Intervertebral fusions in cases of reduced bone density are a tough challenge. From a biomechanical point of view, most current studies have focused on the range of motion or have shown test setups for single-component tests. Definitive setups for biomechanical testing of the primary stability of a 360° fusion using a screw-rod system and cage on osteoporotic spine are missing. The aim of this study was to develop a test stand to provide information about the bone-implant interface under reproducible conditions.

METHODS: After pretesting with artificial bone, functional spine units were tested with 360° fusion in the transforaminal lumbar interbody fusion technique. The movement sequences were conducted in flexion/extension, right and left lateral bending, and right and left axial rotation on a human model with osteopenia or osteoporosis under permanent maximum load with 7.5 N-m.

RESULTS: During the testing of human cadavers, 4 vertebrae were fully tested and were inconspicuous even after radiological and macroscopic examination. One vertebra showed a subsidence of 2 mm, and 1 vertebra had a cage collapsed into the vertebra.

CONCLUSIONS: This setup is suitable for biomechanical testing of cyclical continuous loads on the spine with reduced bone quality or osteoporosis. The embedding method is stable and ensures a purely single-level setup with different trajectories, especially when using the cortical bone trajectory. Optical monitoring provides a very accurate indication of cage movement, which correlates with the macroscopic and radiological results.