Complications at screw removal in slipped capital femoral epiphysis treated by cannulated titanium screws.
Аннотация
INTRODUCTION
Various modes of fixation are proposed for the treatment of slips of the capital femoral epiphysis (SCFE). We describe our experience with the use and removal of a new, cannulated titanium screw (Asnis III, Stryker, Howmedica).
PATIENTS AND
METHODS
Single cannulated titanium screws had been inserted in 101 hips of 65 patients in the 3-year period from 2001 to 2003. These pins were used for in situ fixation of minor chronic slips in 41 patients and for prophylactic fixation on the contralateral side in patients with open physis in 60 patients.
RESULTS
The insertion of these screws was achieved without any real problem. The mean surgical time was 25 min (13-46 min). Problems came up when we started to remove the pins. Hardware retrieval was attempted in 27 patients with 43 pins. The mean surgical time for removal was double the average time of insertion with 51 min (26-107 min). The hexagonal Allen sockets proved to be too weak to overcome the necessary torque for loosening the pin from bone and applying the reverse-cutting-force, necessary to extract the pin. Eleven patients needed extensive chiselling. Two adolescents sustained a subtrochanteric fracture 5 and 7 weeks after hardware removal. Seven pins could not be totally removed.
CONCLUSION
Due to the considerable disadvantages encountered in our series we conclude that Asnis III cannulated screws should be suspended from further use in SCFE.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.