Медицинская библиотека · Practical radiation oncology · 2023

Quantification of MRI Artifacts in Carbon Fiber Reinforced Polyetheretherketone Thoracolumbar Pedicle Screw Constructs prior to Spinal Stereotactic Radiosurgery.

de Almeida Romulo Augusto Andrade, Ghia Amol J, Amini Behrang, Wang Chenyang, Alvarez-Breckenridge Christopher A, Li Jing, Rhines Laurence D, Tom Martin C, North Robert Y, Beckham Thomas H, Tatsui Claudio E

СтатьяПеревод в очереди

Аннотация

PURPOSE: Carbon fiber reinforced polyetheretherketone (CFRP) is a nonmetallic material that is a subject of growing interest in the field of spinal instrumentation manufacturing. The radiolucency and low magnetic susceptibility of CFRP has potential to create less interference with diagnostic imaging compared with titanium implants. However, an objective comparison of the image artifact produced by titanium and CFRP implants has not been described. Spinal oncology, particularly after resection of spinal tumors and at the time of spinal stereotactic radiosurgery planning, relies heavily on imaging interpretation for evaluating resection, adjuvant treatment planning, and surveillance. We present a study comparing measurements of postoperative magnetic resonance imaging artifacts between titanium and CFRP pedicle screw constructs in the setting of separation surgery for metastatic disease.

METHODS AND MATERIALS: The diameter of the signal drop around the screws (pedicle screw artifact) and the diameter of the spinal canal free from artifacts (canal visualization) were measured in consecutive patients who had spinal instrumentation followed by spinal stereotactic radiosurgery in the June 2019 to May 2022 timeframe. The spinal cord presented a shift at the screw level in sagittal images which was also measured (Sagittal Distortion, SagD).

RESULTS

Fifty patients, corresponding to 356 screws and 183 vertebral levels, were evaluated overall. CFRP produced less artifacts in all the 3 parameters compared with titanium: mean pedicle screw artifact (CFRP = 5.8 mm, Ti = 13.2 mm), canal visualization (CFRP = 19.2 mm, Ti = 15.5 mm), and SagD (CFRP = .5 mm, Ti = 1.9 mm), all P < .001. In practice, these findings translate into better-quality magnetic resonance imaging.

CONCLUSIONS

The initial perceived advantages are easier evaluation of postoperative imaging, facilitating radiation treatment planning, recurrence detection, and avoidance in repeating a suboptimal computed tomography myelogram. Further clinical studies analyzing long-term outcomes of patients treated with CFRP implants are necessary.

Для цитирования: de Almeida Romulo Augusto Andrade, Ghia Amol J, Amini Behrang, Wang Chenyang, Alvarez-Breckenridge Christopher A, Li Jing, Rhines Laurence D, Tom Martin C, North Robert Y, Beckham Thomas H, Tatsui Claudio E. Quantification of MRI Artifacts in Carbon Fiber Reinforced Polyetheretherketone Thoracolumbar Pedicle Screw Constructs prior to Spinal Stereotactic Radiosurgery.. Practical radiation oncology. 2023. DOI: 10.1016/j.prro.2023.10.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.