MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/01.bpb.0000060285.16932.7e

Медицинская библиотека · Journal of pediatric orthopedics. Part B · 2003

Guidewire damage during cannulated screw fixation for slipped capital femoral epiphysis.

Robb James E, Annan Iain H, Macnicol Malcolm F

Клинический случайСтатьяПеревод в очереди

Аннотация

Cannulated screw fixation of slipped capital femoral epiphysis is a standard procedure for the mild to moderately displaced slip. We report six cases in which the guidewire was damaged by a cannulated drill. In five the wire broke off within the femur. The broken wire was retrieved in two of the cases but not in the remaining three. The progress of all six patients was unaffected by this complication.

Для цитирования: Robb James E, Annan Iain H, Macnicol Malcolm F. Guidewire damage during cannulated screw fixation for slipped capital femoral epiphysis.. Journal of pediatric orthopedics. Part B. 2003. DOI: 10.1097/01.bpb.0000060285.16932.7e.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.