DOI 10.1097/brs.0000000000002865
Predictive Factors of Postoperative Dysphagia in Single-Level Anterior Cervical Discectomy and Fusion.
Аннотация
STUDY
DESIGN
Retrospective review of prospectively collected data.
OBJECTIVE
To investigate if zero profile devices offer an advantage over traditional plate/cage constructs for dysphagia rates in single level anterior cervical discectomy and fusion (ACDF).
SUMMARY OF BACKGROUND DATA: Dysphagia rates following ACDF have been reported to be as high as 83%, most cases are self-limiting, but chronic dysphagia can continue in up to 35% of patients. Zero profile devices were developed to limit dysphagia, and other plate specific complications, however the literature is currently divided regarding their efficacy.
METHODS
Dysphagia was assessed by swallowing quality of life (SWAL-QOL) scores preoperatively, at 6 weeks and 12 weeks. Patient reported outcome measures (PROMs) including visual analog scale (VAS) and Neck Disability Index (NDI) were collected preoperatively, at 6 weeks and at 6 months. Univariate and multivariate regression analysis was conducted with SWAL-QOL score as the dependent variable.
RESULTS
Sixty-four patients were included, 41 received a zero profile device, and 23 received plate-graft construct. Both groups were similar regarding patient demographics, except operative time, with the zero-profile group having a shorter procedure time than the cage-plate group (44.88 ± 6.54 vs. 54.43 ± 14.71 min, P = 0.001). At all timepoints dysphagia rates were similar between the groups. Regression analysis confirmed preoperative SWAL-QOL and operative time were the only significant variables. PROMs were also similar between groups at all time points, except VAS neck at 6 months, which was lower in the plate-graft group (1.05 ± 1.48 vs. 3.43 ± 3.21, P = 0.007).
CONCLUSION
Operative time and preoperative SWAL-QOL scores are predictive of dysphagia in single level ACDF. Zero profile devices had a significantly shorter operative time, and may provide a benefit in dysphagia rates in this regard.
LEVEL OF EVIDENCE: 3.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.