DOI 10.1016/j.clineuro.2022.107564

Медицинская библиотека · Clinical neurology and neurosurgery · 2022

Factors associated with hardware failure after lateral thoracolumbar fusions - A ten year case series.

Pressman Elliot, Liaw Deborah, Monsour Molly, Wang Christopher P, Gassie Kelly, Alikhani Puya

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

Thoracolumbar lateral interbody fusions (tLLIF) are one tool in the spine surgeon's toolbox to indirectly decompress neuroforamina while also improving segmental lordosis in a biomechanically distinct manner from posterior fusions. When part of a concomitant posterior construct, hardware failure (HF), sometimes requiring revision surgery, can occur. We sought to study the relationship between tLLIF and HF.

METHODS

We conducted a retrospective study on consecutive patents who underwent tLLIF at a single academic center between January 2012 and December 2021 by seven unique neurosurgeons. Patients were excluded if they had no posterior instrumentation within their construct or if they had less than six months of follow-up. Hardware failure was defined as screw breakage or rod fracture seen on postoperative imaging.

RESULTS

232 patients were identified; 6 (2.6 %) developed HF throughout a mean follow-up of 1182 days (range =748-1647 days). Adjacent segment disease was the most common pathology addressed (75 patients (32.3 %)). The amount of posterior instrumentation both in the surgery in question and in the total construct were significantly higher in the HF cohort (4.33 ± 1.52 levels, 5.83 ± 3.36 levels) versus the non-HF cohort (2.08 ± 0.296 levels, p = 0.014; 2.86 ± 0.316 levels, p = 0.003, respectively). The number of interbody devices added in the index surgery and in the entire construct were both significantly higher in the HF cohort (3.33 ± 0.666 interbody devices, 3.33 ± 0.666 devices) than in the non-HF cohort (1.88 ± 0.152 interbody devices, p = 0.002; 2.31 ± 0.158 devices, p = 0.036, respectively). Higher amounts of lateral levels of fusion approached significance for association with HF (HF: 2.67 ± 0.844 levels, no HF: 1.73 ± 1.26 levels, p = 0.076). On multivariate analysis, only the number of interbody devices added in the index surgery was predictive of HF (Odds ratio=2.3, 95 % confidence interval=1.25-4.23, p = 0.007).

CONCLUSION

Greater levels of posterior fusion, and greater numbers of interbody devices in an index surgery and in a construct as a whole, were associated with higher rates of HF in our cohort of patients with tLLIF. Greater numbers of lateral segments fused in this population may also be related to HF.

Для цитирования: Pressman Elliot, Liaw Deborah, Monsour Molly, Wang Christopher P, Gassie Kelly, Alikhani Puya. Factors associated with hardware failure after lateral thoracolumbar fusions - A ten year case series.. Clinical neurology and neurosurgery. 2022. DOI: 10.1016/j.clineuro.2022.107564.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.