Variability in BIA-Derived Muscle Mass Estimates: Device Choice Impacts Diagnostic Classification.
Аннотация
Background/
Objectives
Although discrepancies between bioelectrical impedance analysis (BIA) devices are well documented, their clinical relevance in vulnerable populations remains unclear. This study aims to assess the impact of device choice on muscle mass classification criteria in patients with cancer or obesity and to identify modifiers of device variability.
Methods
BIA data from 224 adults (85 with cancer, 139 with obesity) measured with two segmental multi-frequency devices (seca mBCA 515 and InBody 970) were analyzed. Device differences were assessed using the Wilcoxon signed-rank test and agreement analyses. Differences in classification of body composition cut-offs cited in the GLIM criteria for malnutrition and the ESPEN and EASO criteria for sarcopenic obesity were evaluated using McNemar's test. The impact of disease type, sex, and age on device differences was examined through multivariable models.
Results
Significant device differences were found for all parameters (all p ≤ 0.005). Discrepancies were largest for skeletal muscle mass (kg and %), with effect sizes r > 0.8 and poor agreement (Lin's CCC < 0.90). A significant impact of device choice on muscle mass classification was observed for both cancer and obesity patients (p < 0.001), with seca classifying more patients as having low fat-free mass (50% vs. 20%) and as having a body composition consistent with sarcopenic obesity (90% vs. 50%) than InBody. Discrepancies were more pronounced in cancer patients and females.
Conclusions
Muscle mass assessment by BIA is highly dependent on device choice, potentially leading to clinically relevant discrepancies in classification when rigid cut-offs are applied. An individualized interpretation of BIA data and further validation of prediction equations in disease-specific subpopulations is warranted.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.