Failure of the linkage mechanism in a semi-constrained total elbow arthroplasty is a rare and unpredictable event: a review of seven cases.

Chen Po-An, Chen Alvin Chao-Yu, Cheng Chun-Ying

ОбзорСтатьяПеревод в очереди

Аннотация

PURPOSE: Linked component of total elbow arthroplasty (TEA) consisted of bushing and locking pins. Failure of linked components is a rare complication of TEA. This study aims to investigate the mechanism and consequence of failure of the linkage mechanism in TEA surgeries.

METHODS

Between 2010 and 2021, five patients received revision operation due to linked component failure. Besides, two patients underwent primary operation at another institute were also analyzed due to failure of the linkage mechanism.

RESULTS

All seven patients underwent primary TEA and mean age for primary TEA was 48 (range, 27-62). Two patients had TEA for post-traumatic arthritis, three patients for rheumatoid arthritis, and two patients for comminuted distal humerus fracture. The average time between primary TEA and revision TEA for linked component failure was 13.6 years. Three bushing wear and four locking pin dissociation were diagnosed according to pre-operative radiography. Elbow pain and swelling are the most common clinical symptoms. Severe osteolysis, periprosthetic fracture, and stem loosening were noted in three bushing wear cases. In four dissociation of locking pin cases, breakage of male locking pin phalanges was demonstrated in two patients. For revision procedures, both the locking pins and bushings were replaced. No patients in the study required additional surgery after the revision operation for linked component failure.

CONCLUSION

Osteolysis, component loosening, periprosthetic fracture may be expected after linked component failure. Patients should be regularly followed up from short-term to long-term with radiography. Early diagnosis and intervention with linked component exchange can prevent extensive revision surgery.

Для цитирования: Chen Po-An, Chen Alvin Chao-Yu, Cheng Chun-Ying. Failure of the linkage mechanism in a semi-constrained total elbow arthroplasty is a rare and unpredictable event: a review of seven cases.. International orthopaedics. 2023. DOI: 10.1007/s00264-023-06015-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление