Early failure of bioabsorbable anterior cervical fusion plates: case report and failure analysis.

Brkaric Mario, Baker Kevin C, Israel Raj, Harding Trevor, Montgomery David M, Herkowitz Harry N

Клинический случайСтатьяПеревод в очереди

Аннотация

STUDY

DESIGN

Case report with forensic failure analysis.

OBJECTIVE

To determine the failure modes of 3 explanted 70:30 PLDLA Mystique (Medtronic Sofamor Danek, Memphis, TN) graft containment plates retrieved from revision surgery for early device failure.

SUMMARY OF BACKGROUND DATA: To reduce the problems of stress-shielding and radiopacity associated with metallic systems, bioabsorbable polymers have been used in anterior cervical discectomy and fusion procedures. Degradation of mechanical properties in vivo is a major concern when using bioabsorbable systems. Three of 6 patients who underwent anterior cervical discectomy with instrumented fusion, using Mystique graft containment systems experienced early failure requiring revision to alternate hardware.

METHODS

Devices were retrieved after failure and analyzed by light microscopy and environmental scanning electron microscopy. Simulations were performed with an unused plating system to induce damage for comparison with the retrieved devices. A detailed case review was performed to identify possible sources of extraordinary loading or damage.

RESULTS

One plating system failed at 6 weeks postimplantation due to fatigue fracture of the screws. Crack initiation sites were identified at the interface of the thread root and mold line of the screw. Another plating system failed at 16 weeks postimplantation due to the coalescence of radial microcracking between holes in the plate, leading to catastrophic failure of the plate. The final plating system failed during the implantation surgery, when the screw fractured in torsion.

CONCLUSIONS

Stress concentrations at the screw head-shaft interface and thread-shaft interface reduce the fatigue performance of bioabsorbable screws. Hydrolysis of the polymer may also play a role in the reduction of resistance to crack initiation and propagation.

Для цитирования: Brkaric Mario, Baker Kevin C, Israel Raj, Harding Trevor, Montgomery David M, Herkowitz Harry N. Early failure of bioabsorbable anterior cervical fusion plates: case report and failure analysis.. Journal of spinal disorders & techniques. 2007. DOI: 10.1097/bsd.0b013e318030d2f3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.