DOI 10.1148/radiology.200.2.8685349
Imaging of osseous and cartilaginous intraarticular bodies in the knee: comparison of MR imaging and MR arthrography with CT and CT arthrography in cadavers.
Аннотация
PURPOSE: To compare magnetic resonance (MR) imaging and MR arthrography with computed tomography (CT) and CT arthrography in the detection of intraarticular bodies in the knee.
MATERIALS AND
METHODS
Cuboid (3- or 6-mm-long sides) osseous and cartilaginous bodies were implanted in 16 cadaveric knee specimens. MR imaging was performed with T1-weighted spin-echo (SE), T2-weighted SE, proton-density-weighted SE, gradient recalled acquisition in the steady state (GRASS), and spoiled GRASS sequences. MR arthrography was performed in two phases with saline and 2 mmol/L gadopentetate dimeglumine. CT and CT arthrography were performed in the transaxial plane.
RESULTS
MR arthrography yielded the highest accuracy for the detection of osseous and cartilaginous bodies combined (92%) and was significantly (P < .01) better than MR imaging (57%-70%), CT arthrography (80%), and CT (74%). Accuracy of CT arthrography was significantly better than that of MR imaging and that of CT. Accuracy of saline-enhanced MR arthrography was significantly inferior (P < .001) to that of gadolinium-enhanced MR arthrography.
CONCLUSION
MR arthrography is the best imaging technique for detection of individual intraarticular bodies. CT arthrography is the second most accurate method. Spoiled GRASS and T2-weighted SE sequences are the most accurate at MR imaging. The presence of intraarticular fluid and performance of saline-enhanced MR arthrography improve detectability of intraarticular bodies.