MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.spinee.2022.05.006

Медицинская библиотека · The spine journal : official journal of the North American Spine Society · 2022

Outstanding in vivo mechanical integrity of additively manufactured spinal cages with a novel "honeycomb tree structure" design via guiding bone matrix orientation.

Ishimoto Takuya, Kobayashi Yoshiya, Takahata Masahiko, Ito Manabu, Matsugaki Aira, Takahashi Hiroyuki, Watanabe Ryota, Inoue Takayuki, Matsuzaka Tadaaki, Ozasa Ryosuke, Hanawa Takao, Yokota Katsuhiko, Nakashima Yoshio, Nakano Takayoshi

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Therapeutic devices for spinal disorders, such as spinal fusion cages, must be able to facilitate the maintenance and rapid recovery of spinal function. Therefore, it would be advantageous that future spinal fusion cages facilitate rapid recovery of spinal function without secondary surgery to harvest autologous bone.

PURPOSE: This study investigated a novel spinal cage configuration that achieves in vivo mechanical integrity as a devise/bone complex by inducing bone that mimicked the sound trabecular bone, hierarchically and anisotropically structured trabeculae strengthened with a preferentially oriented extracellular matrix.

STUDY DESIGN/SETTINGS: In vivo animal study.

METHODS

A cage possessing an anisotropic through-pore with a grooved substrate, that we termed "honeycomb tree structure," was designed for guiding bone matrix orientation; it was manufactured using a laser beam powder bed fusion method through an additive manufacturing processes. The newly designed cages were implanted into sheep vertebral bodies for 8 and 16 weeks. An autologous bone was not installed in the newly designed cage. A pull-out test was performed to evaluate the mechanical integrity of the cage/bone interface. Additionally, the preferential orientation of bone matrix consisting of collagen and apatite was determined.

RESULTS

The cage/host bone interface strength assessed by the maximum pull-out load for the novel cage without an autologous bone graft (3360±411 N) was significantly higher than that for the conventional cage using autologous bone (903±188 N) after only 8 weeks post-implantation.

CONCLUSIONS

These results highlight the potential of this novel cage to achieve functional fusion between the cage and host bone. Our study provides insight into the design of highly functional spinal devices based on the anisotropic nature of bone.

CLINICAL SIGNIFICANCE: The sheep spine is similar to the human spine in its stress condition and trabecular bone architecture and is widely recognized as a useful model for the human spine. The present design may be useful as a new spinal device for humans.

Для цитирования: Ishimoto Takuya, Kobayashi Yoshiya, Takahata Masahiko, Ito Manabu, Matsugaki Aira, Takahashi Hiroyuki, Watanabe Ryota, Inoue Takayuki, Matsuzaka Tadaaki, Ozasa Ryosuke, Hanawa Takao, Yokota Katsuhiko, Nakashima Yoshio, Nakano Takayoshi. Outstanding in vivo mechanical integrity of additively manufactured spinal cages with a novel "honeycomb tree structure" design via guiding bone matrix orientation.. The spine journal : official journal of the North American Spine Society. 2022. DOI: 10.1016/j.spinee.2022.05.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.