MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/bcr-2024-264090

Медицинская библиотека · BMJ case reports · 2025

Removal of a 'cold-welded' femoral head lag screw in a nail combination.

Campbell Ryan J, Symes Michael, Sivakumar Brahman, Dragan Zac, Ruff Stephen, Ellis Andrew

СтатьяКлинический случайПеревод в очереди

Аннотация

We present a case of a femoral head lag screw 'cold welded' to an intramedullary nail (IMN) in a female in her 20s during exchange femoral nailing of a subtrochanteric femoral osteotomy non-union. We subsequently developed a reverse spiral-fluted bolt extraction socket, designed to engage the outer wall of a bolt or screw. The lag screw was removed from the Smith and Nephew Meta-Tan IMN, 1 week later, and exchange nailing was successfully completed. This case report describes a technically challenging situation of a cold-welded femoral head lag screw during a complex revision. This report introduces a novel removal device and technique for such cases.

Для цитирования: Campbell Ryan J, Symes Michael, Sivakumar Brahman, Dragan Zac, Ruff Stephen, Ellis Andrew. Removal of a 'cold-welded' femoral head lag screw in a nail combination.. BMJ case reports. 2025. DOI: 10.1136/bcr-2024-264090.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.