Strength and fatigue resistance of thoracolumbar spine implants: an experimental study of selected clinical devices.

Stambough J L, El Khatib F, Genaidy A M, Huston R L

СтатьяПеревод в очереди

Аннотация

Results of strength and endurance tests of several screw/rod thoracolumbar spinal implants are presented. The devices tested were: AcroMed ISOLA, CD systems, Depuy systems, and Synergy systems. The AcroMed ISOLA devices had the greatest stiffness and strength. The Synergy and AcroMed systems had the greatest fatigue resistance. The predominant failure mode was screw bending, although several instances of rod fracture and screw fracture were observed. A number of design improvements are identified, including reducing the number of components, thereby reducing the number of failure mechanisms; avoiding notches and knurled surfaces to provide greater resistance to crack initiation; and strengthening attachments between components, particularly between pedicle screws and horizontal rods.

Для цитирования: Stambough J L, El Khatib F, Genaidy A M, Huston R L. Strength and fatigue resistance of thoracolumbar spine implants: an experimental study of selected clinical devices.. Journal of spinal disorders. 1999.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.