DOI 10.1016/j.wneu.2021.07.013
Biomechanical Evaluation of a Novel S-Type, Dynamic Zero-Profile Cage Design for Anterior Cervical Discectomy and Fusion with Variations in Bone Graft Shape: A Finite Element Analysis.
Аннотация
BACKGROUND
Variations in cage design, material, and graft shape can affect osteointegration and adjacent segment range of motion (ROM) and stress after anterior cervical discectomy and fusion (ACDF) surgery. This study aimed to evaluate the biomechanical properties of a novel dynamic cervical cage design in both titanium (Ti) and polyether ether ketone (PEEK) with variations in bone graft shape using a single level ACDF (FE) model.
METHODS
A 3-dimensional C3-C6 FE model was developed using computed tomography scan data from a healthy male subject. The novel S-shaped dynamic interbody fusion cage with a zero-profile fixation was inserted at the C4-C5 level with 4 different bone graft shapes (square, circular, rectangular, and elliptical). Changes in segmental ROM and maximum von Mises stresses at the fusion and adjacent segments were analyzed.
RESULTS
Both Ti and PEEK cages showed decreased ROM at the fusion and adjacent levels for all shapes of bone graft when compared with the intact spine model. The elliptical graft, for both Ti and PEEK cages, showed a lower percentage of reduction in segmental ROM at the fusion and adjacent levels (0%-5.6%) when compared with other graft shapes (0%-12%). Maximum stresses at the fusion level were lowest in Ti cage with elliptical graft (229.8-347.6 MPa) when compared with other shapes (241.2-476.2 MPa) in flexion, extension, and lateral bending. For the bone graft, maximum stresses were highest on the elliptical-shaped bone graft in flexion and extension in the Ti cage, and in flexion and lateral bending in the PEEK cage.
CONCLUSIONS
Both Ti and PEEK cages showed decreased ROM at the fusion and adjacent levels for all shapes of bone graft when compared with the intact spine model. In the Ti and PEEK dynamic cages, the elliptical shape bone graft showed decreased stress on the cage and increased stress on the bone graft. Further experimental and clinical studies are needed to confirm these encouraging biomechanical results of this novel dynamic, zero-profile fusion device with elliptical bone graft in ACDF surgery.