DOI 10.1016/j.otsr.2021.102954

Медицинская библиотека · Orthopaedics & traumatology, surgery & research : OTSR · 2021

Sagittal reduction of spinal deformity: Superior versus lateral screw-rod connection.

Prost Solène, Pesenti Sébastien, Farah Kaissar, Tropiano Patrick, Fuentes Stéphane, Blondel Benjamin

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

INTRODUCTION

Spinal malalignment can greatly impact a patient's quality of life. Various sagittal parameters are used as realignment goals; however, about 50% of patients end up being under-corrected postoperatively. To improve the correction, prebent rods are available with a radius of curvature corresponding to the patient's "ideal" sagittal alignment. But no studies have been done on how the radius of curvature changes according to the type of connection between the pedicle screws and rods. The goal of this experimental study was to quantify how much prebent rods flatten based on the method used to connect the screw and rod: top-loading screw vs. dome screw with lateral connector.

METHODS

The experiment was done on a material testing system in axial compression on three constructs consisting of two rods secured with top-loading screws and three other constructs consisting of two rods secured with dome screws and lateral connector. The maximum angle of the construct was measured during loading and after removing the load. The primary outcome measure was the mean angle in each construct at each step.

RESULTS

The mean angle of the constructs with top-loading screws when subjected to 500 N load was significantly less than in the constructs with dome screws and lateral connector: 18.6° vs. 24.5° respectively (p<0.0003). The mean angle of the constructs with top-loading screws after removing the load was significantly less than in the constructs with dome screws and lateral connector: 25.7° vs. 32.3° respectively, (p<0.0005).

CONCLUSION

In vitro, top-loading screws produced significantly greater flattening than dome screws with lateral connector. These findings must be confirmed in vivo. Understanding the behavior of rods as a function of the type of screw connection can be an important factor to minimize the risk of under-correction in the sagittal plane.

LEVEL OF EVIDENCE: III.

Для цитирования: Prost Solène, Pesenti Sébastien, Farah Kaissar, Tropiano Patrick, Fuentes Stéphane, Blondel Benjamin. Sagittal reduction of spinal deformity: Superior versus lateral screw-rod connection.. Orthopaedics & traumatology, surgery & research : OTSR. 2021. DOI: 10.1016/j.otsr.2021.102954.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.