MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00256-004-0763-0

Медицинская библиотека · Skeletal radiology · 2004

High-resolution MR imaging of the elbow using a microscopy surface coil and a clinical 1.5 T MR machine: preliminary results.

Yoshioka Hiroshi, Ueno Teruko, Tanaka Toshikazu, Kujiraoka Yuka, Shindo Masashi, Takahashi Nobuyuki, Nishiura Yasumasa, Ochiai Naoyuki, Saida Yukihisa

Клиническое исследованиеКонтролируемое клиническое исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To obtain high-resolution MR images of the elbow using a microscopy surface coil with a 1.5 T clinical machine and to evaluate the feasibility of its use for elbow injuries.

DESIGN AND PATIENTS: Five asymptomatic normal volunteers and 13 patients with elbow pain were prospectively studied with MR imaging using a microscopy surface coil 47 mm in diameter. High-resolution MR images using a microscopy coil were obtained with fast spin echo (FSE) proton density-weighted sequence, gradient recalled echo (GRE) T2*-weighted sequence, and short tau inversion recovery (STIR) sequence, with a 1-2 mm slice thickness, a 50-70 mm field of view, an imaging matrix of 140-224 x 512 using zero fill interpolation, and 2-6 excitations.

RESULTS

High-resolution MR images of normal volunteers using a microscopy coil clearly showed each structure of the medial and lateral collateral ligaments on GRE T2*-weighted images and FSE proton-density weighted images. Partial medial collateral ligament injury, a small avulsion of the medial epicondyle, and osteochondritis dissecans were well demonstrated on high-resolution MR images.

CONCLUSION

High-resolution MR imaging of the elbow using a microscopy surface coil with a 1.5 T clinical machine is a promising method for accurately characterizing the normal anatomy of the elbow and depicting its lesions in detail.

Для цитирования: Yoshioka Hiroshi, Ueno Teruko, Tanaka Toshikazu, Kujiraoka Yuka, Shindo Masashi, Takahashi Nobuyuki, Nishiura Yasumasa, Ochiai Naoyuki, Saida Yukihisa. High-resolution MR imaging of the elbow using a microscopy surface coil and a clinical 1.5 T MR machine: preliminary results.. Skeletal radiology. 2004. DOI: 10.1007/s00256-004-0763-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.