Image-guided placement of occipitocervical instrumentation using a reference arc attached to the headholder.
Аннотация
OBJECTIVE
To develop a safe and accurate method of image-guided placement of instrumentation in the upper cervical spine and occiput in which the reference arc is fixed to the headholder.
METHODS
The authors describe a technique for placing screws at the occipital, C1, and C2 levels using 3-dimensional image guidance in which the reference arc is fixed to the headholder. Technical details are discussed as well as modifications to the technique to maximize navigation accuracy and decrease the need for re-registration. One of 2 paired systems, the BrainLAB Vector Vision system (BrainLAB Inc., Westchester, IL) used in conjunction with the Arcadis Orbic Isocentric C-arm (Siemens Medical Solutions, Erlangen, Germany) or the Stealth Treon system (Medtronic, Littleton, MA) paired with the O-arm (Medtronic), was used for image guidance in this study. A total of 18 patients had 82 screws placed at the occipital, C1, or C2 level using this technique. An independent radiologist interpreted postoperative computed tomographic scans of these patients and graded the screws for bony breach.
RESULTS
No complications resulted from the use of image guidance or from the placement of instrumentation. Postoperative computed tomography revealed 1 screw with a minimal breach of the outer lamina of C2. Another screw was replaced intraoperatively secondary to a minimal bony breach. No other bony breach occurred.
CONCLUSIONS
This technique allows safe and accurate placement of instrumentation in the posterior occipitocervical junction using 3-dimensional image guidance in which the reference arc is attached to the headholder.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.