Biomechanical Analysis of an Interspinous Process Fixation Device with In Situ Shortening Capabilities: Does Spinous Process Compression Improve Segmental Stability?

Wagener Christopher, Gandhi Anup, Ferry Chris, Farmer Sam, DenHaese Ryan

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The objective of this study was to characterize the biomechanical implications of spinous process compression, via in situ shortening of a next-generation interspinous process fixation (ISPF) device, in the context of segmental fusion.

METHODS

Seven lumbar cadaveric spines (L1-L4) were tested. Specimens were first tested in an intact state, followed by iterative instrumentation at L2-3 and subsequent testing. The order followed was 1) stand-alone ISPF (neutral height); 2) stand-alone ISPF (shortened in situ from neutral height; shortened); 3) lateral lumbar interbody fusion (LLIF) + ISPF (neutral); 4) LLIF + ISPF (shortened); 5) LLIF + unilateral pedicle screw fixation; 6) LLIF + bilateral pedicle screw fixation. A 7.5-Nm moment was applied in flexion/extension, lateral bending, and axial rotation via a kinematic test frame. Segmental range of motion (ROM) and lordosis were measured for all constructs. Comparative analysis was performed.

RESULTS

Statistically significant flexion/extension ROM reductions: all constructs versus intact condition (P < 0.01); LLIF + ISPF (neutral and shortened) versus stand-alone ISPF (neutral and shortened) (P < 0.01); LLIF + USPF versus ISPF (neutral) (P = 0.049); bilateral pedicle screw fixation (BPSF) versus stand-alone ISPF (neutral and shortened) (P < 0.01); LLIF + BPSF versus LLIF + unilateral pedicle screw fixation (UPSF) (P < 0.01). Significant lateral bending ROM reductions: LLIF + ISPF (neutral and shortened) versus intact condition and stand-alone ISPF (neutral) (P < 0.01); LLIF + UPSF versus intact condition and stand-alone ISPF (neutral and shortened) (P < 0.01); LLIF + BPSF versus intact condition and all constructs (P < 0.01). Significant axial rotation ROM reductions: LLIF + ISPF (shortened) and LLIF + UPSF versus intact condition and stand-alone ISPF (neutral) (P ≤ 0.01); LLIF + BPSF versus intact condition and all constructs (P ≤ 0.04).

CONCLUSIONS

In situ shortening of an adjustable ISPF device may support increased segmental stabilization compared with static ISPF.

Для цитирования: Wagener Christopher, Gandhi Anup, Ferry Chris, Farmer Sam, DenHaese Ryan. Biomechanical Analysis of an Interspinous Process Fixation Device with In Situ Shortening Capabilities: Does Spinous Process Compression Improve Segmental Stability?. World neurosurgery. 2020. DOI: 10.1016/j.wneu.2020.08.203.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.