DOI 10.1097/00007632-200002010-00008

Медицинская библиотека · Spine · 2000

Biomechanical studies of a dynamized anterior thoracolumbar implant.

Hitchon P W, Goel V K, Rogge T, Grosland N M, Sairyo K, Torner J

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

DESIGN

An in vitro investigation into the biomechanical properties of a dynamized anterolateral compression implant that allows controlled subsidence.

OBJECTIVES

To determine the extent to which both modes of the anterolateral compression implant (controlled collapsing and rigid) are able to reestablish the stability of the lumbar spine after L4 corpectomy.

SUMMARY OF BACKGROUND DATA: Over time, anterior and posterior spinal implants have been associated with progressive angulation, and occasionally implant failure and breakage. To circumvent this occurrence and provide better graft loading, dynamized or collapsing devices for clinical use have been developed.

METHODS

Eight fresh calf spines (L1-L6) were placed in a biomechanical testing frame. Pure moments of 6 Nm were loaded onto the intact spine in six directions: flexion, extension, right and left lateral bending, and right and left axial rotation. A total L4 corpectomy then was performed, and the defect grafted with a wooden dowel. Loading was repeated after the specimens were stabilized using the two modes of the anterolateral compression implant in succession.

RESULTS

The results showed that both modes of the implant (the rigid mode in particular) restore the stiffness of the unstable spine to normal levels of flexion, extension, and right and left lateral bending, even to levels exceeding normal. These devices, however, fall short of achieving normal stability in right and left axial rotation.

CONCLUSION

In the cadaveric calf spine after L4 corpectomy, restoration of stability with a dynamized anterior spinal implant is possible in flexion, extension, and right and left lateral bending, but not in axial rotation.

Для цитирования: Hitchon P W, Goel V K, Rogge T, Grosland N M, Sairyo K, Torner J. Biomechanical studies of a dynamized anterior thoracolumbar implant.. Spine. 2000. DOI: 10.1097/00007632-200002010-00008.