DOI 10.1007/s00586-024-08410-7

Медицинская библиотека · European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2024

Repeatability and intra and interrater reproducibility of the DIPA-S eHealth© capture and analysis system for clinical assessment of scoliosis.

Navarro Isis Juliene Rodrigues Leite, Silva Marcelle Guimarães, Dos Santos Ingrid Cláudia Pereira, Candotti Cláudia Tarragô

СтатьяПеревод в очереди

Аннотация

PURPOSE: To assess the repeatability, intra and interrater reproducibility of the DIPA-S eHealth© system for capturing and measuring clinical variables of scoliosis, including frontal trunk imbalance (FTI), sagittal trunk imbalance (STI), and angle of trunk rotation (ATR).

METHODS

Patients were photographed using the DIPA-S eHealth Capture© mobile application by family members, physiotherapists, or surgeons. Three photos were taken in each position: standing in the frontal and sagittal planes and in the axial plane in forward bending position of the trunk. The photos were analyzed by three independent evaluators using the DIPA-S eHealth Analysis© software. For repeatability, each photo was analyzed twice consecutively by the three evaluators. For intrarater reproducibility, only one evaluator reanalyzed the first photo from each plane with a 5-day interval. For interrater reproducibility, the three evaluators analyzed the first photo from each plane. The intraclass correlation coefficient (ICC), standard error of measurement (SEM), and minimal detectable change (MDC) were used (p < 0.05).

RESULTS

The sample comprised 262 images (n = 30). Repeatability was excellent for all three evaluators in all three planes (ICC 0.94 to 1.00). Intrarater reproducibility was excellent in all three planes (ICC 0.88 to 0.99). Interrater analysis was excellent for the frontal and axial planes (ICC 0.98 and 0.93), respectively. However, it was weak in the sagittal plane ICC 0.32. The SEM ranged from 0.02 to 17.6 and MDC from 0.03 to 34.6.

CONCLUSION

The DIPA-S eHealth© Capture and Analysis system demonstrates reproducibility for use in the clinical assessment of scoliosis through teleconsultations.

Для цитирования: Navarro Isis Juliene Rodrigues Leite, Silva Marcelle Guimarães, Dos Santos Ingrid Cláudia Pereira, Candotti Cláudia Tarragô. Repeatability and intra and interrater reproducibility of the DIPA-S eHealth© capture and analysis system for clinical assessment of scoliosis.. European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2024. DOI: 10.1007/s00586-024-08410-7.