Ventricular arrhythmia detection with a wearable ring-type photoplethysmography sensor: A feasibility study.

van Steijn Nicole J, Pepplinkhuizen Shari, Postema Pieter G, Knops Reinoud E, Winter Michiel M

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Ventricular arrhythmias (VAs) are most frequently the onset of sudden cardiac death, and early detection is key for survival. Limited studies suggest that wrist-worn photoplethysmography (PPG) sensors can detect VA. Ring-type wearable PPG sensors may be more accurate; however, they have not yet been investigated.

OBJECTIVE

We aimed to show the sensitivity of a ring-type wearable PPG sensor to detect induced VA.

METHODS

Patients undergoing a procedure involving VA induction were included. The CART-I was positioned on the finger during the procedure, and PPG and reference electrocardiogram data were collected. A loss of cardiac output was defined as a flatline, or a decrease in amplitude of ≥50%. Sensitivity was calculated using a contingency table. Bland-Altman analysis and intraclass correlation coefficients were used to assess the agreement between PPG- and electrocardiogram-derived arrhythmia durations.

RESULTS

Thirty-seven patients were enrolled, of whom 25 were included: defibrillation threshold testing (n = 16), ventricular tachycardia (VT) ablation (n = 8), and electrophysiological study (n = 1). A total of 67 VA episodes, of which 27 were ventricular fibrillation (VF) and 40 were VT episodes, were recorded. A loss of cardiac output was observed on the tachograms of all VF episodes and 36 VT episodes (90%). The sensitivity was 94% (VF, 100%; VT, 90%). The mean difference in VA duration was 0.1 seconds (95% limits of agreement -1.4 to 1.7), with an intraclass correlation coefficient of 0.998 (95% confidence interval, 0.997-0.999), indicating excellent agreement.

CONCLUSION

This study suggests that a ring-type wearable device can detect, with high sensitivity, changes in arterial pulse flow during induced VA.

Для цитирования: van Steijn Nicole J, Pepplinkhuizen Shari, Postema Pieter G, Knops Reinoud E, Winter Michiel M. Ventricular arrhythmia detection with a wearable ring-type photoplethysmography sensor: A feasibility study.. Heart rhythm. 2025. DOI: 10.1016/j.hrthm.2025.06.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.