DOI 10.1016/j.acmx.2016.10.005

Медицинская библиотека · Archivos de cardiologia de Mexico · 2016

[Increase in the effectiveness of identifying peaks and feet of the photoplethysmographic pulse to be reconstructed it using adaptive filtering].

Becerra-Luna Brayans, Martínez-Memije Raúl, Cartas-Rosado Raúl, Infante-Vázquez Oscar

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To improve the identification of peaks and feet in photoplethysmographic (PPG) pulses deformed by myokinetic noise, through the implementation of a modified fingertip and applying adaptive filtering.

METHOD: PPG signals were recordedfrom 10 healthy volunteers using two photoplethysmography systems placed on the index finger of each hand. Recordings lasted three minutes andwere done as follows: during the first minute, both handswere at rest, and for the lasting two minutes only the left hand was allowed to make quasi-periodicmovementsin order to add myokinetic noise. Two methodologies were employed to process the signals off-line. One consisted on using an adaptive filter based onthe Least Mean Square (LMS) algorithm, and the other includeda preprocessing stage in addition to the same LMS filter. Both filtering methods were compared and the one with the lowest error was chosen to assess the improvement in the identification of peaks and feet from PPG pulses.

RESULTS

Average percentage errorsobtained wereof 22.94% with the first filtering methodology, and 3.72% withthe second one. On identifying peaks and feet from PPG pulsesbefore filtering, error percentages obtained were of 24.26% and 48.39%, respectively, and once filtered error percentageslowered to 2.02% for peaks and 3.77% for feet.

CONCLUSIONS

The attenuation of myokinetic noise in PPG pulses through LMS filtering, plusa preprocessing stage, allows increasingthe effectiveness onthe identification of peaks and feet from PPG pulses, which are of great importance for medical assessment.

Для цитирования: Becerra-Luna Brayans, Martínez-Memije Raúl, Cartas-Rosado Raúl, Infante-Vázquez Oscar. [Increase in the effectiveness of identifying peaks and feet of the photoplethysmographic pulse to be reconstructed it using adaptive filtering].. Archivos de cardiologia de Mexico. 2016. DOI: 10.1016/j.acmx.2016.10.005.