MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00590-024-03981-z

Медицинская библиотека · European journal of orthopaedic surgery & traumatology : orthopedie traumatologie · 2024

Optimizing hook implantation angle of the clavicular hook plate: a cadaveric study.

Saengsirinavin Pamok, Iamsirikulmit Chumnarn, Piyapittayanun Peerapong, Phiphobmongkol Vajarin, Jongthanakamol Thanawan, Ratanalekha Rosarin

СтатьяПеревод в очереди

Аннотация

PURPOSE: While Clavicle hook plates have demonstrated favorable results regarding bone and shoulder function, their design can potentially lead to complications due to pressure concentration at the plate's tip. This study aims to investigate the impact of different hook implantation angles on the contact surface area between the hook plate and acromion, with the goal of minimizing mismatch and maximizing contact surface area.

METHODS

Twenty soft shoulder cadavers were included in the study, and the contact surface area of the hook plate was measured in different positions based on the hook implantation angle.

RESULTS

The results showed variations in compatibility, width, and length of the contact surface area depending on the hook implantation angle and the medial or lateral row placement. The lateral row generally demonstrated superior compatibility (84.0% vs 46.67%, p-value < 0.001), with a broader contact area (3.55 ± 0.08 mm vs 3.09 ± 0.10 mm, p-value = 0.004) and a longer contact area (7.36 ± 0.19 mm vs 5.10 ± 0.23 mm, p-value < 0.001) at specific angles. A detailed analysis of the lateral position revealed that the zero angle of implantation resulted in the greatest contact surface area, measuring 3.91 ± 0.70 mm in width (p value = 0.083) and 8.85 ± 1.24 mm in length (p value < 0.001).

CONCLUSION

Placing the hook laterally and at the zero position according to the hook implantation angle can maximize contact surface area, may reduce stress concentration, and minimize complications in hook plate fixation. Further research and consideration of anatomical variations are warranted to refine the placement technique and enhance patient outcomes.

LEVEL OF EVIDENCE: Level V evidence.

Для цитирования: Saengsirinavin Pamok, Iamsirikulmit Chumnarn, Piyapittayanun Peerapong, Phiphobmongkol Vajarin, Jongthanakamol Thanawan, Ratanalekha Rosarin. Optimizing hook implantation angle of the clavicular hook plate: a cadaveric study.. European journal of orthopaedic surgery & traumatology : orthopedie traumatologie. 2024. DOI: 10.1007/s00590-024-03981-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.