Biomechanical analysis and review of lateral lumbar fusion constructs.

Cappuccino Andrew, Cornwall G Bryan, Turner Alexander W L, Fogel Guy R, Duong Huy T, Kim Kee D, Brodke Darrel S

СтатьяПеревод в очереди

Аннотация

STUDY

DESIGN

Biomechanical study and the review of literature on lumbar interbody fusion constructs.

OBJECTIVE

To demonstrate the comparative stabilizing effects of lateral interbody fusion with various supplemental internal fixation options.

SUMMARY OF BACKGROUND DATA: Lumbar interbody fusion procedures are regularly performed using anterior, posterior, and more recently, lateral approaches. The biomechanical profile of each is determined by the extent of resection of local supportive structures, implant size and orientation, and the type of supplemental internal fixation used.

METHODS

Pure moment flexibility testing was performed using a custom-built 6 degree-of-freedom system to apply a moment of ±7.5 Nm in each motion plane, while motion segment kinematics were evaluated using an optoelectronic motion system. Constructs tested included the intact spine, stand-alone extreme lateral interbody implant, interbody implant with lateral plate, unilateral and bilateral pedicle screw fixation. These results were evaluated against those from literature-reported biomechanical studies of other lumbar interbody constructs.

RESULTS

All conditions demonstrated a statistically significant reduction in range of motion (ROM) as a percentage of intact. In flexion-extension, ROM was 31.6% stand-alone, 32.5% lateral fixation, and 20.4% and 13.0% unilateral and bilateral pedicle screw fixation, respectively. In lateral bending, the trend was similar with greater reduction with lateral fixation than in flexion-extension; ROM was 32.5% stand-alone, 15.9% lateral fixation, and 21.6% and 14.4% unilateral and bilateral pedicle screw fixation. ROM was greatest in axial rotation; 69.4% stand-alone, 53.4% lateral fixation, and 51.3% and 41.7% unilateral and bilateral pedicle screw fixation, respectively.

CONCLUSION

The extreme lateral interbody construct provided the largest stand-alone reduction in ROM compared with literature-reported ALIF and TLIF constructs. Supplemental bilateral pedicle screw-based fixation provided the overall greatest reduction in ROM, similar among all interbody approach techniques. Lateral fixation and unilateral pedicle screw fixation provided intermediate reductions in ROM. Clinically, surgeons may evaluate these comparative results to choose fixation options commensurate with the stability requirements of individual patients.

Для цитирования: Cappuccino Andrew, Cornwall G Bryan, Turner Alexander W L, Fogel Guy R, Duong Huy T, Kim Kee D, Brodke Darrel S. Biomechanical analysis and review of lateral lumbar fusion constructs.. Spine. 2010. DOI: 10.1097/brs.0b013e318202308b.