Performance and reliability assessment of a lower dose, task-based scoliosis radiography protocol in pediatric patients.

Al-Dasuqi Khalid, Taylor Erin, Ehrlich Lauren, Cooperman Daniel, Socci Adrienne, Tuason Dominick, Hoerner Matthew, Staib Lawrence, Silva Cicero T

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Follow-up scoliosis radiographs are performed to assess the degree of spinal curvature and skeletal maturity, which can be done at lower radiation exposures than those in standard-dose radiography.

OBJECTIVE

Describe and evaluate a protocol that reduced the radiation in follow-up frontal-view scoliosis radiographs.

MATERIALS AND

METHODS

We implemented a postero-anterior lower dose modified-technique for scoliosis radiography with task-based definition of adequate image quality and use of technique charts based on target exposure index and patient's height and weight. We subsequently retrospectively evaluated 40 consecutive patients who underwent a follow-up radiograph using the modified-technique after an initial standard-technique radiograph. We evaluated comparisons of proportions for subjective assessment with chi-squared tests, and agreements of reader's scores with intraclass correlation coefficients and Bland-Altman plots. We determined incident air kerma, exposure index, deviation index/standard deviation, dose-area product (DAP), and effective dose for each radiograph. We set statistical significance at P<0.05.

RESULTS

Forty patients (65% female), aged 4-17 years. Median effective dose was reduced from 39 to 10 µSv (P<0.001), incident air kerma from 139 to 29 µSv (P<0.001), and DAP from 266 to 55 mGy*cm2 (P<0.001). All modified-technique parameters were rated with a mean score of acceptable or above. All modified-technique measurements obtained inter- and intra-observer correlation coefficient agreements of 0.86 ("Good") or greater.

CONCLUSION

Substantial dose reduction on follow-up scoliosis imaging with existing radiography units is achievable through task-based definition of adequate image quality and tailoring of radiation to each patient's height and weight, while still allowing for reliable assessment and reproducible measurements.

Для цитирования: Al-Dasuqi Khalid, Taylor Erin, Ehrlich Lauren, Cooperman Daniel, Socci Adrienne, Tuason Dominick, Hoerner Matthew, Staib Lawrence, Silva Cicero T. Performance and reliability assessment of a lower dose, task-based scoliosis radiography protocol in pediatric patients.. Pediatric radiology. 2023. DOI: 10.1007/s00247-023-05812-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.