Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices.

Tsuji Hironori, Misawa Haruo, Takigawa Tomoyuki, Tetsunaga Tomoko, Yamane Kentaro, Oda Yoshiaki, Ozaki Toshifumi

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Deep tendon reflexes are one of the main components of the clinical nervous system examinations. These assessments are inexpensive and quick. However, evaluation can be subjective and qualitative. This study aimed to objectively evaluate hyperreflexia of the patellar tendon reflex using portable mechanomyography (MMG) and electromyography (EMG) devices. This study included 10 preoperative patients (20 legs) who had a pathology that could cause bilateral patellar tendon hyperreflexia and 12 healthy volunteers (24 legs) with no prior history of neurological disorders. We attached MMG/EMG sensors onto the quadriceps and tapped the patellar tendon with maximal and constant force. Our results showed a significantly high amplitude of the root mean square (RMS) and low frequency of the mean power frequency (MPF) in the rectus femoris, vastus medialis, and vastus lateralis muscles in both EMG and MMG with both maximal and constant force. Especially in the patients with cervical and thoracic myelopathy, the receiver operating characteristic (ROC) curve for diagnosing hyperreflexia of the patellar tendon showed a moderate to very high area under the curve for all EMG-RMS, EMG-MPF, MMG-RMS, and MMG-MPF values. The use of EMG and MMG for objectively quantifying the patellar tendon reflex is simple and desirable for future clinical applications and could help diagnose neurological disorders.

Для цитирования: Tsuji Hironori, Misawa Haruo, Takigawa Tomoyuki, Tetsunaga Tomoko, Yamane Kentaro, Oda Yoshiaki, Ozaki Toshifumi. Quantification of patellar tendon reflex using portable mechanomyography and electromyography devices.. Scientific reports. 2021. DOI: 10.1038/s41598-021-81874-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.