DOI 10.1016/j.spinee.2006.07.004

Медицинская библиотека · The spine journal : official journal of the North American Spine Society · 2007

Static versus dynamic plating for multilevel anterior cervical discectomy and fusion.

DuBois Christian M, Bolt Patrick M, Todd Andrew G, Gupta Purnendu, Wetzel F Todd, Phillips Frank M

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Dynamic anterior plates have been popularized to promote cervical spine fusion by allowing controlled settling, thereby promoting load sharing across the construct. To date these proposed benefits have been largely theoretical and there are no studies confirming any benefits over more traditional static plates.

PURPOSE: To compare the clinical and radiographic outcomes of patients undergoing an instrumented multilevel anterior cervical discectomy and fusion (ACDF) with either a static or dynamic plate design.

STUDY

DESIGN

A retrospective clinical and radiographic study.

PATIENT SAMPLE: From 1997 to 2002, 52 patients with either radiculopathy or myelopathy underwent two- or three-level ACDF with either static or dynamic plate fixation.

OUTCOME MEASURES: Functional outcome, fusion status, radiographic measurements.

METHODS

A statically locked plating system was used in 21 patients, and a dynamic plating system was used in 31 patients. Functional outcome, fusion status, plate migration, settling, and adjacent-level disc space impingement were evaluated.

RESULTS

Clinical outcome was found to be similar between the statically and dynamically plated groups. Eighty-four percent of patients in both groups experienced good or excellent results at final follow-up. We observed a higher rate of nonunion in patients treated with a dynamic plate (16% [5 of 31]) compared with a rate of 5% (1 of 21) in those patients treated with a static plate (p=.05). Settling of the construct and plate migration was similar between the study groups at all time points.

CONCLUSIONS

This study failed to confirm our hypothesis that a dynamic plate (that allows angular motion between the screws and plate) confers any clinical or radiographic advantage over earlier design static plates. A higher rate of nonunion was actually seen in the dynamically plated patients; however, clinical results were similar between the two groups.

Для цитирования: DuBois Christian M, Bolt Patrick M, Todd Andrew G, Gupta Purnendu, Wetzel F Todd, Phillips Frank M. Static versus dynamic plating for multilevel anterior cervical discectomy and fusion.. The spine journal : official journal of the North American Spine Society. 2007. DOI: 10.1016/j.spinee.2006.07.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.