MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.clinbiomech.2003.12.003

Медицинская библиотека · Clinical biomechanics (Bristol, Avon) · 2004

Selection of fixation devices in proximal femur rotational osteotomy: clinical complications and finite element analysis.

Chen Weng-Pin, Tai Ching-Lung, Shih Chun-Hsiung, Hsieh Pang-Hsin, Leou Menq-Chiang, Lee Mel S

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To compare the postoperative stability between large cancellous screws fixation and dynamic hip screws fixation in transtrochanteric rotational osteotomy.

DESIGN

Finite element analysis was designed to validate the clinical outcomes of fixation failure in transtrochanteric rotational osteotomy.

BACKGROUND

From 1997 to 2000, transtrochanteric rotational osteotomy had been done in 20 osteonecrotic hips. Fixation with large cancellous screws as suggested by Sugioka had been used in the first five hips. Unfortunately, fixation failures were encountered in all of the five cases. The fixation mode was then shifted to a combination of plate and screws, no fixation failure had been observed postoperatively. A finite element analysis was designed to investigate the causes of fixation failure.

METHODS

Finite element models for four dynamic hip screw instrumented femora and four large screws instrumented femora simulating four different levels of bone cut were created. The von Mises stress distributions of each model were analyzed and compared for a loading condition simulating single-legged stance.

RESULTS

Analysis results confirmed our clinical observation that fixation with screws had higher stress concentration on the proximal femur. Improper bone cut, especially those that are too distal from the intertrochanteric line, had higher risks of fixation failure.

CONCLUSIONS

Transtrochanteric rotational osteotomy is a technically demanding procedure such that surgical principles should be abided carefully to avoid catastrophic complications.

RELEVANCE: Finite element analysis results demonstrated that dynamic hip screw fixation provides better stability and prevents fixation failure that corresponds with the clinical observation.

Для цитирования: Chen Weng-Pin, Tai Ching-Lung, Shih Chun-Hsiung, Hsieh Pang-Hsin, Leou Menq-Chiang, Lee Mel S. Selection of fixation devices in proximal femur rotational osteotomy: clinical complications and finite element analysis.. Clinical biomechanics (Bristol, Avon). 2004. DOI: 10.1016/j.clinbiomech.2003.12.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.