DOI 10.1007/s13239-023-00693-z
Evaluation of Wrist-Worn Photoplethysmography Trackers with an Electrocardiogram in Patients with Ischemic Heart Disease: A Validation Study.
Аннотация
PURPOSE: Photoplethysmography measurement of heart rate with wrist-worn trackers has been introduced in healthy individuals. However, additional consideration is necessary for patients with ischemic heart disease, and the available evidence is limited. The study aims to evaluate the validity and reliability of heart rate measures by a wrist-worn photoplethysmography (PPG) tracker compared to an electrocardiogram (ECG) during incremental treadmill exercise among patients with ischemic heart disease.
METHODS
Fifty-one participants performed the standard incremental treadmill exercise in a controlled laboratory setting with 12-lead ECG attached to the patient's body and wearing wrist-worn PPG trackers.
RESULTS
At each stage, the absolute percentage error of the PPG was within 10% of the standard acceptable range. Further analysis using a linear mixed model, which accounts for individual variations, revealed that PPG yielded the best performance at the baseline low-intensity exercise. As the stages progressed, heart rate validity decreased but was regained during recovery. The reliability was moderate to excellent.
CONCLUSIONS
Low-cost trackers AMAZFIT Cor and Bip validity and reliability were within acceptable ranges, especially during low-intensity exercise among patients with ischemic heart disease recovering from cardiac procedures. Though using the tracker as part of the diagnosis tool still requires more supporting studies, it can potentially be used as a self-monitoring tool with precautions.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.