Interference Screw Versus Suture Endobutton Fixation of a Fiber-Reinforced Meniscus Replacement Device in a Human Cadaveric Knee Model.
Аннотация
BACKGROUND
Meniscal lesions represent one of the most common intra-articular knee injuries. Meniscus replacement devices are needed to restore load distribution and knee stability after meniscectomy. Fixation of these devices is crucial to the generation of hoop stresses and the distribution of loads in the joint.
PURPOSE: To evaluate 2 different fixation techniques (suture endobutton and interference screw) for implantation of a novel meniscus device.
STUDY
DESIGN
Controlled laboratory study.
METHODS
In 7 human cadaveric knees (aged 17-61 years), 1 anterior and 2 potential posterior tunnel locations were investigated, and both fixation techniques were tested in each tunnel. The native meniscus roots, devices fixed with a suture endobutton, and devices fixed with an interference screw were gripped with cryoclamps, and tibias were drilled and loaded into a custom jig. Samples were preloaded, preconditioned, loaded for 500 cycles (50-150 N), and tested in tension until failure.
RESULTS
For all 3 tunnels, suture fixation resulted in greater elongation (54.1%-150.7% greater; P < .05) during cyclic loading than interference screw fixation, which approximated the native roots. Both fixation techniques displayed ultimate tensile loads in the same range as native roots. However, stiffness of the suture fixation groups (36.5-41.6 N/mm) was only 28% to 37% of that of the interference screw fixation groups (98.7-131.6 N/mm), which had values approaching those of the native roots (anterior: 175.4 ± 24.2 N/mm; posterior: 157.6 ± 22.9 N/mm).
CONCLUSION
Interference screw fixation was found to be superior to suture fixation with regard to elongation and stiffness, a finding that should be considered in the design and implantation of novel meniscus replacement devices.
CLINICAL RELEVANCE: With the emergence of various devices for total meniscus replacement, the establishment of fixation strategies is crucial for the generation of tensile hoop stresses and the efficacy of these approaches.