DOI 10.1016/j.foot.2017.11.001
A locking compression plate versus the gold-standard non-locking plate with lag screw for first metatarsophalangeal fusion: A biomechanical comparison.
Аннотация
INTRODUCTION
The treatment of end-stage first metatarso-phalangeal joint (MTP) arthritis has been arthrodesis. A dorsal non-locking plate with a lag screw has been the standard traditional fixation method. This study compares the biomechanical strength of a locking compression plate (LCP) with and without internal compression versus this known gold standard.
METHODS
In group 1, six matched pairs of cadaver great toes were used to compare the standard non-locking dorsal plate and 3.5mm lag screw to an anatomic locking compression plate in which a lag screw was utilized rather than the internal compression features of the plate. In group 2, another six matched pairs of cadaver great toes were used to compare the gold standard to the locking compression plate, utilizing the plate's internal compression feature instead of a lag screw. A material testing system (MTS) machine applied loads to the MTP joints and measured displacement and stiffness of the constructs. The stiffness of the constructs (Young's modulus) was calculated from the force-displacement curves, and the displacement was measured.
RESULTS
The locking compression plate group that used the compression features of the plate, without the lag screw, had less joint displacement and higher stiffness than control (p<0.05). The same plating construct in which a lag screw was used rather than internal compression of the plate was found to be stiffer than the control (p<0.05), but displacement was not statistically significant.
DISCUSSION: The results suggest that a locking compression plate alone provides the stiffest construct for a first MTP joint fusion.