Comparison of 3-dimensional spinal reconstruction accuracy: biplanar radiographs with EOS versus computed tomography.

Glaser Diana A, Doan Josh, Newton Peter O

Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

DESIGN

Experimental study for systematic evaluation of 3-dimensional (3D) reconstructions from low-dose digital stereoradiography.

OBJECTIVE

To assess the accuracy of EOS (EOS Imaging, Paris, France) 3-dimensional (3D) reconstructions compared with 3D computed tomography (CT) and the effect spine positioning within the EOS unit has on reconstruction accuracy.

SUMMARY OF BACKGROUND DATA: Scoliosis is a 3D deformity, but 3D morphological analyses are still rare. A new low-dose radiation digital stereoradiography system (EOS) was previously evaluated for intra/interobserver variability, but data are limited for 3D reconstruction accuracy.

METHODS

Three synthetic scoliotic phantoms (T1-pelvis) were scanned in upright position at 0°, ±5°, and ±10° of axial rotation within EOS and in supine position using CT. Three-dimensional EOS reconstructions were superimposed on corresponding 3D computed tomographic reconstructions. Shape, position, and orientation accuracy were assessed for each vertebra and the entire spine. Additional routine planer clinical deformity measurements were compared: Cobb angle, kyphosis, lordosis, and pelvic incidence.

RESULTS

Mean EOS vertebral body shape accuracy was 1.1 ± 0.2 mm (maximum 4.7 mm), with 95% confidence interval of 1.7 mm. Different anatomical vertebral regions were modeled well with root-mean-square (RMS) values from 1.2 to 1.6 mm. Position and orientation accuracy of each vertebra were high: RMS offset was 1.2 mm (maximum 3.7 mm) and RMS axial rotation was 1.9° (maximum 5.8°). There was no significant difference in each of the analyzed parameters (P > 0.05) associated with varying the rotational position of the phantoms in EOS machine. Planer measurements accuracy was less than 1° mean difference for pelvic incidence, Cobb angle (mean 1.6°/maximum 3.9°), and sagittal kyphosis (mean less than 1°, maximum 4.9°).

CONCLUSION

The EOS image acquisition and reconstruction software provides accurate 3D spinal representations of scoliotic spinal deformities. The results of this study provide spinal deformity surgeons evidence pertaining to this new upright 3D imaging technology that may aid in the clinical diagnosis and decision making for patients with scoliosis.

Для цитирования: Glaser Diana A, Doan Josh, Newton Peter O. Comparison of 3-dimensional spinal reconstruction accuracy: biplanar radiographs with EOS versus computed tomography.. Spine. 2012. DOI: 10.1097/brs.0b013e3182518a15.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.