Comparison of clinically used bilayer collagen membrane and trilayer collagen prototype fixation stability in chondral defects at the talus - An experimental human specimen study.

Richter Martinus, Milstrey Alexander, Zech Stefan, Evers Julia, Grueter Angelika, Raschke Michael J, Ochman Sabine

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The purpose of this human specimen experimental study was to compare the fixation stability of clinically used bilayer collagen membrane with fibrin glue with trilayer collagen prototype without fibrin glue in chondral defects at the medial or lateral talar shoulder (both matrices from Geistlich Pharma AG, Wollhusen, Switzerland).

METHODS

Eleven human specimens were used. The membranes were implanted in standardized chondral defects at the medial and lateral talar shoulder randomized. All tests were performed in load-control 15 kg. Range of motion ROM of each ankle was examined individually before testing. The average ROM was 10° dorsiflexion range 0°-20° and 30° plantarflexion range 20°-45°. 1,000 testing cycles with the defined ROM were performed. Two independent investigators, blinded to membrane and fixation type, visually assessed the membrane fixation integrity for peripheral detachment, area of defect uncovered, membrane constitution and delamination.

RESULTS

The clinically used bilayer collagen membrane plus fibrin glue showed higher fixation stability than the trilayer prototype (all p < 0.05). No significant differences occurred between medial and lateral talar shoulder location (all p > 0.05).

CONCLUSIONS

The fixation stability of the trilayer collagen prototype without fibrin glue is lower than of the clinically used bilayer membrane with fibrin glue in chondral defects at the medial and lateral talar shoulder in an experimental human specimen test. Clinical use of trilayer collagen prototype without fibrin glue has to be validated by clinical testing to evaluate if the lower stability of fixation is still sufficient.

Для цитирования: Richter Martinus, Milstrey Alexander, Zech Stefan, Evers Julia, Grueter Angelika, Raschke Michael J, Ochman Sabine. Comparison of clinically used bilayer collagen membrane and trilayer collagen prototype fixation stability in chondral defects at the talus - An experimental human specimen study.. Foot and ankle surgery : official journal of the European Society of Foot and Ankle Surgeons. 2020. DOI: 10.1016/j.fas.2020.01.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.