m+Medelevator.RU
← Поиск публикаций
European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2021

Metallosis is commonly associated with magnetically controlled growing rods; results from an independent multicentre explant database.

СтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Rushton Paul R P, Smith Simon L, Fender David, Bowey Andrew J, Gibson Michael J, Joyce Thomas J

Аннотация

PURPOSE: Determine the incidence of metallosis around MAGEC rods.

METHODS: A multicentre explant database was searched to identify cases with complete intraoperative findings at rod removal. Surgeons removing rods detailed the presence or absence of tissue metallosis associated with rods. More recently surgeons measured the 'length' of tissue metallosis. Prior to rod disassembly, the majority underwent testing with an external remote controller (ERC). The impact of clinical and explant variables on metallosis was assessed.

RESULTS: Sixty-six cases were identified. Mean age at insertion was 8.1 ± 2.3 years with mean duration of implantation 37.6 ± 15.1 months. Tissue metallosis was noted at revision surgery in 52/66 cases (79%). Metallosis was noted more commonly when rods were removed during fusion surgery than rod removal/exchange (97% vs. 58% (p =  < 0.01)). The mass at insertion was greater in cases with metallosis (25.9 ± 7.8 kg vs. 21.1 ± 6.2 kg, p = 0.04). Length of tissue metallosis was reported for 45 rods, median 9 cm (range 1-25). Metallosis was noted in 43/59 (73%) rods that produced no force and 22/30 (73%) rods that produced some force on ERC activation (p = 0.96). Wear debris was found within the actuator in all rods, and all but 3 rods had damaged O-rings.

CONCLUSION: MAGEC rods are associated with tissue metallosis in the majority of cases. It is seen with functional rods as well as failed rods and appears related to wear debris within the actuator and high rates of O-ring failure. Until the implications of metal debris in children are known, we urge caution with the use of this implant.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты