INTRODUCTION

The fractures around the supramalleolar region are difficult to manage and deformities may occur following insufficient or inadequate treatment attempts. The treatment of those deformities is even more challenging. The available options for surgical fixation have inherent problems considering the soft tissue problems, infection and compliance of the patients. In this study, we report the preliminary outcome of new IMN system in use of supramalleolar deformities.

PATIENTS AND

METHODS

Three patients with supramalleolar deformities were recruited for this study. Two patients had periarticular distal tibia malunions and one patient had nonunion in same region. All patients were operated with using newly designed intramedullary nail system with a unique distal interlocking bolt screw. The outcome for each patient was followed-up and evaluated with using AOFAS score.

RESULTS

The patients were followed up for 3 years, 3 months, and 6 months, respectively. The AOFAS scores were 36, 33, and 21 preoperatively; and 75, 68, and 75 postoperatively in Patients 1, 2 and 3, respectively.

CONCLUSION

In our patient series, adequate fixation following correction of the deformity was seen. Although the number of the cases are limited this study provides encouraging results regarding the outcome of new IMN system in use of supramalleolar deformities.

Для цитирования: Küçükdurmaz Fatih, Sağlam Necdet, Kurtulmuş Tuhan, Akpınar Fuat. A novel intramedullary nail for use in the treatment of supramalleolar malunion and nonunion: A preliminary report of three cases.. Acta orthopaedica et traumatologica turcica. 2016. DOI: 10.1016/j.aott.2014.11.001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.