m+Medelevator.RU
← Поиск публикаций
Clinical biomechanics (Bristol, Avon) · 2012

Kinematics of a selectively constrained radiolucent anterior lumbar disc: comparisons to hybrid and circumferential fusion.

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Daftari Tapan K, Chinthakunta Suresh R, Ingalhalikar Aditya, Gudipally Manasa, Hussain Mir, Khalil Saif

Аннотация

BACKGROUND: Despite encouraging clinical outcomes of one-level total disc replacements reported in literature, there is no compelling evidence regarding the stability following two-level disc replacement and hybrid constructs. The current study is aimed at evaluating the multidirectional kinematics of a two-level disc arthroplasty and hybrid construct with disc replacement adjacent to rigid circumferential fusion, compared to two-level fusion using a novel selectively constrained radiolucent anterior lumbar disc.

METHODS: Nine osteoligamentous lumbosacral spines (L1-S1) were tested in the following sequence: 1) Intact; 2) One-level disc replacement; 3) Hybrid; 4) Two-level disc replacement; and 5) Two-level fusion. Range of motion (at both implanted and adjacent level), and center of rotation in sagittal plane were recorded and calculated.

FINDINGS: At the level of implantation, motion was restored when one-level disc replacement was used but tended to decrease with two-level disc arthroplasty. The findings also revealed that both one-level and two-level disc replacement and hybrid constructs did not significantly change adjacent level kinematics compared to the intact condition, whereas the two-level fusion construct demonstrated a significant increase in flexibility at the adjacent level. The location of center of rotation in the sagittal plane at L4-L5 for the one-level disc replacement construct was similar to that of the intact condition.

INTERPRETATION: The one-level disc arthroplasty tended to mimic a motion profile similar to the intact spine. However, the two-level disc replacement construct tended to reduce motion and clinical stability of a two-level disc arthroplasty requires additional investigation. Hybrid constructs may be used as a surgical alternative for treating two-level lumbar degenerative disc disease.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты