Effect of collimator choice on quantitative assessment of cardiac iodine 123 MIBG uptake.

Inoue Yusuke, Suzuki Akira, Shirouzu Ichiro, Machida Toru, Yoshizawa Yasunori, Akita Fumihide, Ohnishi Satoshi, Yoshikawa Kohki, Ohtomo Kuni

Сравнительное исследованиеОценочное исследованиеСтатьяВалидационное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Quantitative accuracy in iodine 123 studies may be impaired by septal penetration. We evaluated the effect of collimator choice on estimation of the heart-to-mediastinum (H/M) ratio in cardiac I-123 metaiodobenzylguanidine (MIBG) imaging.

METHODS AND

RESULTS

A low-energy high-resolution (LEHR) collimator, special LEHR (SLEHR) collimator, and medium-energy (ME) collimator were used. In experiments in which a phantom of simple geometry was used, the use of the LEHR collimator provided the lowest contrast accuracy, suggesting the effect of septal penetration. Thoracic phantom studies demonstrated contamination of heart and mediastinum counts by lung and liver activities, which was greatest with the LEHR collimator and least with the ME collimator. In 8 patients anterior chest views were acquired successively with the three collimators after I-123 MIBG injection. H/M ratios were significantly higher with the SLEHR collimator than with the LEHR collimator and were still higher with the ME collimator. The difference in H/M ratios between the LEHR and ME collimators showed a high positive correlation with the lung-to-mediastinum ratio.

CONCLUSIONS

Collimator choice substantially influences estimation of the H/M ratios in cardiac I-123 MIBG imaging. The use of an ME collimator provides high quantitative accuracy and may enhance reliability in the evaluation of cardiac sympathetic nerve function.

Для цитирования: Inoue Yusuke, Suzuki Akira, Shirouzu Ichiro, Machida Toru, Yoshizawa Yasunori, Akita Fumihide, Ohnishi Satoshi, Yoshikawa Kohki, Ohtomo Kuni. Effect of collimator choice on quantitative assessment of cardiac iodine 123 MIBG uptake.. Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology. 2003. DOI: 10.1016/s1071-3581(03)00652-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.