[Measurement of femoral neck anteversion of developmental dislocation of hip in children by 3D printing technique].
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Cao Jin, Hu Shan-Shan, Zheng Hua-Jiang, Zhang Nan, Zhong Zhao-Ping
Аннотация
OBJECTIVE: To explore the method and accuracy of measuring the femoral neck anteversion in children with developmental dislocation of the hip by using 3D printing technology, so as to find out the method of measuring the femoral neck anteversion accurately and guide the formulation of the operation plan.
METHODS: From June 2016 to September 2018, 17 patients with unilateral developmental dislocation of the hip were enrolled in the study, including 2 males and 15 females, aged 2 to 13 (5.47±0.81) years old, 11 on the left and 6 on the right. The methods of CT, 3D printing and intraoperative anteversion of femoral neck were used to measure the anteversion of femoral neck respectively. The intraoperative measurement was used as the standard reference value forstatistical analysis, and the accuracy of the first two methods was compared.
RESULTS: The average value of CT was (36.00±1.66)°, the average value of model method was(43.91±1.62)°, and the average value of intraoperative method was(44.21±1.62)°. There were significant differences in CT measurement, model measurement and intraoperative measurement(P<0.05). There was no significant difference between model method and intraoperative method (P>0.05).
CONCLUSION: Compared with the traditional CT method, the 3D printing model method is simpler, more accurate and more repeatable, and can simulate the operation before operation. It is an ideal method to measure the femoral neck anteversion.
METHODS: From June 2016 to September 2018, 17 patients with unilateral developmental dislocation of the hip were enrolled in the study, including 2 males and 15 females, aged 2 to 13 (5.47±0.81) years old, 11 on the left and 6 on the right. The methods of CT, 3D printing and intraoperative anteversion of femoral neck were used to measure the anteversion of femoral neck respectively. The intraoperative measurement was used as the standard reference value forstatistical analysis, and the accuracy of the first two methods was compared.
RESULTS: The average value of CT was (36.00±1.66)°, the average value of model method was(43.91±1.62)°, and the average value of intraoperative method was(44.21±1.62)°. There were significant differences in CT measurement, model measurement and intraoperative measurement(P<0.05). There was no significant difference between model method and intraoperative method (P>0.05).
CONCLUSION: Compared with the traditional CT method, the 3D printing model method is simpler, more accurate and more repeatable, and can simulate the operation before operation. It is an ideal method to measure the femoral neck anteversion.