m+Medelevator.RU
← Поиск публикаций
Archives of orthopaedic and trauma surgery · 2017

Is the novel olecranon tension plate a valid alternative to tension band wiring of olecranon fractures? A biomechanical study on cadaver bones.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Gruszka Dominik, Arand Charlotte, Greenfield Julia, Nowak Tobias E, Kuechle Raphael, Kuhn Sebastian, Rommens Pol M

Аннотация

INTRODUCTION: Tension band wiring (TBW) used for olecranon fractures is afflicted with a high rate of complications. The aim of this study was to compare the stability of the TBW versus an alternative, novel low-profile olecranon tension plate (OTP) with angular stable screws in a simulated complex fracture model.

MATERIALS AND METHODS: Nine fresh-frozen pairs of cadaver proximal ulnae with an oblique osteotomy and an additional wedge fragment simulating an unstable fracture were tested. The TBW and OTP were implanted pairwise. The elbow motion was simulated in a single-muscle model and ranged from full extension to 90° of flexion. The pulling force of the triceps tendon ranged from 50 to 200 N. The displacement of the fracture fragments was measured with a video motion analysis system over 304 cycles. Data were assessed statistically using the Wilcoxon signed-rank test.

RESULTS: The cyclic loading tests showed mean loosening of the fracture fragments at the articular surface of 0.56 mm using TBW (SD 0.65) and 0.15 mm for OTP (SD 0.39). There was no statistical significance in loosening between the two constructs (p = 0.31). No plate breakage or screw loosening occurred.

CONCLUSIONS: The low-profile OTP, using the principles of pre-tensioning, lag, cortical, and angular stable screws together, demonstrated similar stability after olecranon fracture fixation when compared with the TBW technique and a lag screw in the olecranon fracture model with a third wedge fragment.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты