MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1121/1.2735811

Медицинская библиотека · The Journal of the Acoustical Society of America · 2007

The dependence of time-domain speed-of-sound measurements on center frequency, bandwidth, and transit-time marker in human calcaneus in vitro.

Wear Keith A

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Time-domain speed-of-sound (SOS) measurements in calcaneus are effective predictors of osteoporotic fracture risk. High attenuation and dispersion in bone, however, produce severe distortion of transmitted pulses that leads to ambiguity of time-domain SOS measurements. An equation to predict the effects of system parameters (center frequency and bandwidth), algorithm parameters (pulse arrival-time marker), and bone properties (attenuation coefficient and thickness) on time-domain SOS estimates is derived for media with attenuation that varies linearly with frequency. The equation is validated using data from a bone-mimicking phantom and from 30 human calcaneus samples in vitro. The data suggest that the effects of dispersion are small compared with the effects of frequency-dependent attenuation. The equation can be used to retroactively compensate data. System-related variations in SOS are shown to decrease as the pulse-arrival-time marker is moved toward the pulse center. Therefore, compared with other time-domain measures of SOS, group velocity exhibits the minimum system dependence.

Для цитирования: Wear Keith A. The dependence of time-domain speed-of-sound measurements on center frequency, bandwidth, and transit-time marker in human calcaneus in vitro.. The Journal of the Acoustical Society of America. 2007. DOI: 10.1121/1.2735811.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.